Assessing agreement of blood pressure-measuring devices in Tibetan areas of China: a systematic review
  1. Cuomu Mingji1,2,
  2. Igho J Onakpoya3,
  3. Carl J Heneghan3,
  4. Alison M Ward3
  1. 1Tibetan Medical College, Lhasa, Tibet
  2. 2Austrian Academy of Science, Vienna, Austria
  3. 3Nuffield Department of Primary Care Health Sciences, Centre for Evidence-Based Medicine, University of Oxford, Oxford, UK
  1. Correspondence to Dr Igho Onakpoya, Nuffield Department of Primary Care Health Sciences, Centre for Evidence-Based Medicine, University of Oxford, Gibson Building, Radcliffe Observatory Quarter, Oxford OX2 6GG, UK; igho.onakpoya{at}phc.ox.ac.uk

Abstract

Background The validity of blood pressure (BP)-measuring tools at very high altitudes is uncertain. Therefore, the objective of this review was to examine the degree of agreement of BP-measuring devices in Tibet.

Methods We conducted electronic searches in Medline, Embase, Cinahl, Cochrane Library, Global Health Library and the ISI Web of Science. Randomised and observational studies were considered for inclusion. The methodological characteristics of included studies were assessed using the Quality Assessment of Diagnostic Accuracy Studies-2 criteria. Our primary outcome was the difference in mean BP measurements between the new device and the gold standard.

Results We identified three eligible studies, out of which two with a total of 162 participants were included. The studies differed in their methodology. One study reported significantly higher systolic blood pressure (SBP) measurement with electronic sphygmomanometer (Omron) compared with mercury sphygmomanometer (mean difference 5.8±4.7 mm Hg; p<0.001), with no significant difference in diastolic blood pressure (DBP) measurement (0.4±3.9 mm Hg; p=0.23). The second study reported mean differences of 1.0±5.9 mm Hg and −3.1±4.6 mm Hg for SBP and DBP, respectively.

Conclusion The limited evidence from published studies suggests that automated (Omron) BP monitors show a high degree of agreement for DBP when compared against mercury sphygmomanometer at high altitudes. However, the degree of such agreement for SBP is not consistent. Few studies assessing the validity of automated BP monitors at high altitudes have been conducted, and they differ in design and methodology. Further research assessing the suitability of BP-measuring instruments at high altitudes is therefore warranted.

  • EBM
  • HYPERTENSION

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Contributors CM and IJO were involved with abstract screening, data extraction, data analysis and interpretation, and codrafting of the review. CJH and AMW were involved with data analysis and interpretation, and codrafting of the review.

  • Funding CJH receives payment for running educational courses at the University of Oxford and University of Oxford ISIS consulting services for external teaching and training. He also receives royalties for books (Evidence Based Toolkit series by Blackwell BMJ Books).

  • Competing interests None

  • Provenance and peer review Not commissioned; externally peer reviewed.

Linked Articles

  • Editorial
    Willem J Verberk Stephan Mieke
  • Correction
    BMJ Publishing Group Ltd, British Cardiovascular Society and Asia Pacific Heart Association