A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial

Lancet. 2008 Mar 15;371(9616):899-907. doi: 10.1016/S0140-6736(08)60415-8.

Abstract

Background: A fully bioabsorbable drug-eluting coronary stent that scaffolds the vessel wall when needed and then disappears once the acute recoil and constrictive remodelling processes have subsided has theoretical advantages. The bioasorbable everolimus-eluting stent (BVS) has a backbone of poly-L-lactic acid that provides the support and a coating of poly-D,L-lactic acid that contains and controls the release of the antiproliferative agent everolimus. We assessed the feasibility and safety of this BVS stent.

Methods: In this prospective, open-label study we enrolled 30 patients who had either stable, unstable, or silent ischaemia and a single de-novo lesion that was suitable for treatment with a single 3.0 x 12 mm or 3.0 x 18 mm stent. Patients were enrolled from four academic hospitals in Auckland, Rotterdam, Krakow, and Skejby. The composite endpoint was cardiac death, myocardial infarction, and ischaemia-driven target lesion revascularisation. Angiographic endpoints were available for 26 patients and intravascular-ultrasound endpoints for 24 patients. Clinical endpoints were assessed in all 30 patients at 6 and 12 months. In a subset of 13 patients, optical coherence tomography was undertaken at baseline and follow-up. Analysis was by intention to treat. This study is registered with ClinicalTrials.gov, number NCT00300131.

Findings: Procedural success was 100% (30/30 patients), and device success 94% (29/31 attempts at implantation of the stent). At 1 year, the rate of major adverse cardiac events was 3.3%, with only one patient having a non-Q wave myocardial infarction and no target lesion revascularisations. No late stent thromboses were recorded. At 6-month follow-up, the angiographic in-stent late loss was 0.44 (0.35) mm and was mainly due to a mild reduction of the stent area (-11.8%) as measured by intravascular ultrasound. The neointimal area was small (0.30 [SD 0.44] mm2), with a minimal area obstruction of 5.5%.

Interpretation: This study shows the feasibility of implantation of the bioabsorbable everolimus-eluting stent, with an acceptable in-stent late loss, minimal intrastent neointimal hyperplasia, and a low stent area obstruction.

Funding: Abbott Vascular.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants
  • Angioplasty, Balloon, Coronary
  • Coronary Disease / diagnostic imaging
  • Coronary Disease / therapy*
  • Coronary Restenosis
  • Drug-Eluting Stents*
  • Everolimus
  • Female
  • Humans
  • Immunosuppressive Agents*
  • Male
  • Middle Aged
  • Prospective Studies
  • Safety
  • Sirolimus / analogs & derivatives*
  • Ultrasonography

Substances

  • Immunosuppressive Agents
  • Everolimus
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT00300131